Pressure gauge calibration applying 0-A-0 pressurization to reference gauge

Authors

  • Hiroaki Kajikawa National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST)
  • Tokihiko Kobata

DOI:

https://doi.org/10.21014/acta_imeko.v5i1.319

Abstract

A calibration method is proposed to improve pressure calibration that uses a pressure gauge as the reference device. The reference gauge is pressurized by a 0-A-0 pressurization procedure, whereas the test gauge can be pressurized through various procedures. The calibration results with this method were consistent with those calibrated against a pressure balance. The method is expected to help develop low-cost pressure calibration systems that are more precise.

Author Biography

Hiroaki Kajikawa, National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST)

Senior researcher

Downloads

Published

2016-04-29