Traceability and measurement uncertainty of non-removable field flowmeters using clamp-on ultrasonic flowmeters as reference

Authors

  • Catarina Ferreira Simões National Laboratory for Civil Engineering
  • Álvaro Silva Ribeiro National Laboratory for Civil Engineering
  • Maria do Céu Almeida National Laboratory for Civil Engineering
  • Dídia Covas CERIS, Instituto Superior Técnico, Universidade de Lisboa

DOI:

https://doi.org/10.21014/actaimeko.v13i3.1777

Keywords:

Flow measurement, clamp-on ultrasonic flowmeter, measurement uncertainty, traceability, measurement accuracy

Abstract

This study examines the traceability and measurement uncertainty of in situ hydraulic calibration of flowmeters using clamp-on ultrasonic flowmeters as reference. The procedure compares the equipment readings with the reference ones. Measurement uncertainty evaluation uses Guide to the expression of Uncertainty in Measurement (GUM) formulation, considering the linearity conditions of the mathematical models applied. Experimental values are used to test the procedure and its suitability for actual cases where the expected accuracy needs to be achieved.