Reduction of gravity effect on the results of low-frequency accelerometer calibration

Authors

  • G. P. Ripper Division of Acoustics and Vibration Metrology - DIAVI INMETRO
  • C. D. Ferreira Vibration Laboratory - LAVIB INMETRO
  • R. S. Dias Vibration Laboratory - LAVIB INMETRO
  • G. B. Micheli Vibration Laboratory - LAVIB INMETRO

DOI:

https://doi.org/10.21014/acta_imeko.v9i5.1002

Abstract

This paper describes a study on the possible sources of systematic errors during the calibration of accelerometers at low-frequencies. This study was carried out on a primary calibration system that uses an air-bearing vibration exciter APS Dynamics 129 and applying the sine-approximation method. Tests performed and actions taken to reduce the effect on experimental results are presented.