Validation of combinatorial evaluation of strain-gauge amplifier linearity
DOI:
https://doi.org/10.21014/acta_imeko.v9i5.970Abstract
This paper describes a validation of a combinatorial calibration technique based calibration procedure for strain-gauge amplifier calibration. The deviation between two strain-gauge amplifiers at calibration using the combinatorial technique is compared to the deviation between the same two amplifiers at calibration on a force calibration machine as a transducer-amplifier chain under unchanged conditions, serving as an amplifier linearity comparator. This enables the validation to be confirmed with a suitable expanded measurement uncertainty of 3 nV/V.Downloads
Published
2020-12-31
Issue
Section
Conference Submissions
License
Authors who publish with this journal agree to the following terms:
- Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution License that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.
- Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.