Characterization of a liquid bath for accurate thermometer calibration and detailed uncertainty estimation

Authors

  • Oqab AlOtaibi National Measurement and Calibration Center (NMCC), Saudi Standards, Metrology and Quality Organization (SASO), Riyadh, Kingdom of Saudi Arabia
  • Ismail A. AlFaleh National Measurement and Calibration Center (NMCC), Saudi Standards, Metrology and Quality Organization (SASO), Riyadh, Kingdom of Saudi Arabia
  • Adel B. Shehata National Measurement and Calibration Center (NMCC), Saudi Standards, Metrology and Quality Organization (SASO), Riyadh, Kingdom of Saudi Arabia

DOI:

https://doi.org/10.21014/actaimeko.v15i4.2191

Keywords:

liquid bath, uniformity, stability, thermometer calibration, uncertainty

Abstract

The reliable characterization of liquid baths is essential for ensuring traceable and accurate temperature calibrations in metrology laboratories. In this study, a liquid water bath with a copper equalizing block of 12 cm length was characterized at three calibration points of 5 °C, 29.7 °C, and 80 °C. The bath performance was assessed through drift evaluation, stability, axial and radial uniformity, and loading influence using two 25 Ohm (Ω) standard platinum resistance thermometers (SPRTs). The results obtained showed excellent drift performance within ± 1 millikelvins (mK) over 12 hours at 20 °C. The instability at the three calibration points remained below 1 mK. Axial and radial uniformity values did not exceed 0.9 m 2.7 mK, respectively, and the loading effect produced temperature deviations of less than 3 mK. A detailed uncertainty analysis was carried out based on the requirements of the Guide to the expression of uncertainty in measurement (ISO GUM), considering all relevant contributions from the SPRTs and the liquid bath. The combined standard uncertainties of the bath characterization were 2.70, 3.17, and 3.87 mK at 5 °C, 29.7 °C, and 80 °C, respectively, all well below the 10 mK benchmark typically required for precision calibration baths. After the liquid bath characterization, calibration of a platinum resistance thermometer (PRT) by comparison against a 100-Ω SPRT was performed at 5 °C, 29.7 °C, and 80 °C. The associated expanded uncertainties of calibration were 0.052 °C, 0.050 °C, and 0.043 °C, respectively, including the bath characterization uncertainty. These findings confirm that the characterized liquid bath provides a stable and homogeneous thermal environment suitable for high-accuracy thermometer calibrations.

Downloads

Published

2026-11-17

Issue

Section

Research Papers